Specifications
  Fluke Ti25 Fluke Ti10
.Imaging Performance
.Thermal: Field of View (FOV) 23° horizontal x 17° vertical
....Minimum focus distance Thermal lens: 15 cm (approx. 6 in), Visible (visual) light lens: 46 cm (approx. 18 in)
....Thermal sensitivity (NETD) ≤ 0.1 °C at 30 °C (100 mK) ≤ 0.2 °C at 30 °C (200 mK)
....Minimum span (Auto/Manual) 5 °C/2.5 °C 10 °C/5 °C
....Focus Manual
....Detector size 160 x 120
.Visual: On camera operating modes Picture-in-Picture (Blending is user selectable
between MAX, MID and MIN) and full
screen IR (Blending is user selectable
between MAX, MID and MIN)
Full Picture-in-Picture and full screen IR
....Visual (Visible light) camera 640 x 480 pixels, full color
.Temperature measurement
.Temperature Range -20 °C to +350 °C (-4 °F to +662 °F), 2 ranges -20 °C to +250 °C (-4 °F to +482 °F)
.Accuracy ± 2 °C or 2 % (whichever is greater) ± 5 °C or 5 % (whichever is greater)
.Measurement modes Center point and hot and cold markers Center point
.On-screen emissivity correction Yes No
.Image presentation
.Digital display 9.1 cm (3.6 in) diagonal landscape color VGA (640 x 480) LCD
.LCD backlight Selectable bright or auto
.Pallettes Ironbow, blue-red, high contrast
amber, hot metal, grey
Ironbow, blue-red, high contrast, grey
.Image and data storage
.Fully radiometric Yes
.Storage medium 2 GB SD card stores up to 3000 .bmp IR images or 1200 .IS2 IR-Fusion images
.File formats supported Exportable to JPEG, BMP, GIF, PNG, TIFF, WMF, EXIF, and EMF
.Voice memo recorder (voice annotation) Yes No
.Software SmartView; Full analysis and reporting software included
Controls and adjustments
.Set-up controls Date/time, °C/°F, language, emissivity, hot spot
and cold spot on image
Date/time, °C/°F, language
.Language selection English, German, French, Spanish, Portuguese, Italian, Swedish, Finnish, Russian, Czech,
Polish, Turkish, Simplified Chinese, Traditional Chinese, Korean, Japanese
.Image control Smooth auto scaling and manual scaling
.On-screen indicators Battery status, real time clock and center point temperature, range and
span indication and high and low alarm settings
.Power
.Battery type Internal rechargeable batter (included)
.Battery operating time 3 to 4 hours continuous operation
.Battery charging 2 hours with AC charge or DC charger (charges battery while operating)
.AC operation AC adapter/charger 110/230 VAC, 50/60 Hz
.Power saving Automatic shutdown and sleep modes (user specified)
.Environmental and mechanical design
.Operating temperature -10 °C to +50 °C (+14 °F to +122 °F)
.Storage temperature -20 °C to +50 °C (-4 °F to +122 °F)
.Relative humidity Operating and storage 10 % to 90 %, non-condensing
.Water and dust resistant IP54
.Two meter (6.5 feet) drop test Yes
.Protective lens cover Yes
.Weight (including battery) 1.2 kg (2.6 lbs)
.Imager size (HxWxD) 267 mm x 127 mm x 152 mm (10.2 in x 5.0 in x 6.0 in)
.Other
.Warranty Two-years
.EN 61010-1 2nd edition and EN61326-1 Yes