Specifications |
||||
| Fluke Ti55FT | Fluke Ti50FT | Fluke Ti45FT | Fluke Ti40FT | |
| .Imaging Performance | ||||
| .Field of view (FOV) | 23° horizontal x 17° vertical | |||
| .Spatial resolution (IFOV) | 1.30 mrad | 2.6 mrad | ||
| .Min focus distance | 0.15 m (5.9 in) | |||
| .Thermal sensitivity (NETD) | ≤ 0.05 °C at 30 °C (50 mK) | ≤ 0.07 °C at 30 °C (70 mK) | ≤ 0.08 °C at 30 °C (80 mK) | ≤ 0.09 °C at 30 °C (90 mK) |
|
.Detector
data acquisition/ .image frequency |
60 Hz/60Hz | 30 Hz/30 Hz | ||
| .Focus | SmartFocus; one finger continuous focus (manual) | |||
| .IR digital zoom | 2x, 4x, 8x | 2x | 2x | - |
| .Detector type |
320 x 240 Focal Plane Array, Vanadium
Oxide (VOx) Uncooled Microbolometer |
160 x 120 Focal Plane Array, Vanadium
Oxide (VOx) Uncooled Microbolometer |
||
| .Spectral band | 8 µm to 14 µm | |||
| .Digital image enhancement | Automatic full-time enhanced | |||
|
.On
camera IR-Fusion .operating modes |
Full thermal, full visual light or blended thermal-visual images. Picture-in-Picture | |||
| .Visible light camera | 1280 x 1024 pixels, full color | |||
| Visible light digital zoom | 2x, 4x | 2x | 2x | - |
| .Temperature measurement | ||||
|
.Calibrated .temperature range |
.-20
°C to 600 °C (-4 °F to .1112 °F) in three ranges .Range one = -20 °C to .100 °C (-4 °F to 212 °F) .Range two = -20 °C to .350 °C (32 °F to 662 °F) .Range three = 250 °C to .600 °C (482 °F to 1112 °F) |
.-20
°C to 350 °C (-4 °F to .662 °F) in two ranges .Range one = -20 °C to .100 °C (-4 °F to 212 °F) .Range two = -20 °C to .350 °C (32 °F to 662 °F) |
.-20
°C to 600 °C (-4 °F to .1112 °F) in three ranges .Range one = -20 °C to .100 °C (-4 °F to 212 °F) .Range two = -20 °C to .350 °C (32 °F to 662 °F) .Range three = 250 °C to .600 °C (482 °F to 1112 °F) |
.-20
°C to 350 °C (-4 °F to .662 °F) in two ranges .Range one = -20 °C to .100 °C (-4 °F to 212 °F) .Range two = -20 °C to .350 °C (32 °F to 662 °F) |
| .Optional-high temperature |
Up to 1200 °C (2192 °F) |
- | ||
|
.Range
four = 500 °C to .1200 °C (932 °F to 2192 °F) |
- | |||
| .Accuracy | ± 2 °C or 2 % (whichever is greater) | |||
| .Measurement modes |
.Centerpoint,
center box .(area min/max, average), .moveable spots/boxes, .user defined field/text .annotations, automatic hot .and cold point detection, .visible color alarm above .and below |
.Centerpoint,
center box .(area min/max, average) |
.Centerpoint,
center box .(area min/max, average), .moveable spots/boxes, .user defined field/text .annotations, automatic hot .and cold point detection, .visible color alarm above .and below |
.Centerpoint,
center box .(area min/max, average) |
| .Emissivity correction | 0.1 to 1.0 (0.01 increments) | |||
| .Image presentation | ||||
| .Digital display | 0.13 m (5 in) large high-resolution digital display | |||
| .LCD backlight | Sunlight readable color LCD | |||
| .Video output | RS 170 EIA/NTSC or CCIR/PAL composite video | |||
| .Palettes | Grayscale, grayscale inverted, blue red, high contrast, hot metal, ironbow, amber, amber inverted | |||
| .Optical Lens | ||||
| .54 mm telephoto lens |
High precision
Germanium lens |
|||
| Spatial resolution (IFOV): 0.47 mrad | Spatial resolution (IFOV): 0.94 mrad | |||
| Min focus distance: 0.6 m (1.97 ft) | ||||
| .Other | ||||
| .Operating temperature | -10 °C to +50 °C (14 °F to 122 °F) | |||
| .Weight (including batteries) | 1.95 kg (4.3 lb) | |||
| .Camera size (HxWxD) | 162 mm x 262 mm x 101 mm (6.5 in x 10.5 in x 4.0 in) | |||
| .Warranty | Two-years | |||